Agilent Technologies 8510C Manual de servicio Pagina 116

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 222
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 115
3-46 8510XF Network Analyzer Systems
Operation
Operation Using a Wafer Probe Station
Operation Using a Wafer Probe Station
System Configuration For on-wafer measurements, it is usually best to remove the network
analyzer from the instrument rack, and place it on a table adjacent to the
probe station. This allows for easy viewing of the analyzer display, and easy
access to the analyzer’s front panel controls.
The test heads are placed on X-Y positioners on the probe station. Locations
on a probe station are conventionally referred to in terms of compass points;
accordingly, the left test head is placed on the “west” positioner, and the
right test head is placed on the “east” positioner.
Figure 3-23 Wafer Probe Configuration
Vista de pagina 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 221 222

Comentarios a estos manuales

Sin comentarios