8510XF Network Analyzer Systems 3-17
Operation
Calibration Types
One Path 2-Port This calibration is for reflection and transmission measurements, in one
direction only (with Port 1 defined as the source). To perform this calibration
using the 1.0 mm calibration kit, select:
[CAL] {CAL1 or CAL2} {ONE PATH 2-PORT}
This calls up the following menu:
Figure 3-6 One Path 2-Port Menu & Equivalent GPIB Commands
Notes on the Standards {REFLECT’N} These are the same standards that are used for the S11
and S22 1-port calibrations, except that in this case both
ports are calibrated. The reflection standards are sorted
into three categories:
See “Open/Short Standards” on page 3-21
See “Shorts Standards” on page 3-22
See “Load/Short Standards” on page 3-23
{TRANS-
MISSION}
See “One-Path Transmission Standards” on page 3-24.
{ISOLATION} See “One-Path Isolation Standard” on page 3-25.