3-16 8510XF Network Analyzer Systems
Operation
Calibration Types
S11 1-Port This calibration is for reflection measurements using the left test port (within
those operational limits, however, it is a thorough and accurate calibration).
To perform this calibration using the 1.0 mm calibration kit, select:
[CAL] {CAL1 or CAL2} {S11 1-PORT}
This calls up the following menu:
Figure 3-5 S11 1-Port Menu & Equivalent GPIB Commands
For notes on the three types of standards involved in this calibration, see the
following:
“Open/Short Standards” on page 3-21
“Shorts Standards” on page 3-22
“Load/Short Standards” on page 3-23
S22 1-Port This calibration is for reflection measurements using the right test port
(within those operational limits, however, it is a thorough and accurate
calibration). To perform this calibration using the 1.0 mm calibration kit,
select:
[CAL] {CAL1 or CAL2} {S22 1-PORT}
This calls up a menu which is identical to the one shown in Figure 3-5
above, except that it is identified as S22 rather than S11.