Chapter 3: Testing Performance
Offset Accuracy Test
21
is set up as follows and then click OK:
Source = Channel 1
Measurement area = Entire Display
4
Set the channel 1 offset value to 400.0 mV. This can be done using the front
panel control or:
a Pull down the Setup menu and select Channel 1 or click the Channel 1 setup icon.
b Change the vertical sensitivity (Scale) of channel 1 to 10 mV/div.
c Click the Offset control arrows to change the offset value or click on the offset
value and enter 400.0 mV in the dialog box.
5 Set the Cal Out voltage (V
Cal Out
) to +400.0 mV as follows:
a Select Utilities > Calibration Output.
b Change the Signal Output function to DC (top left corner).
c Set the Level to 400.0 mV.
d Click Close.
6 Press [Clear Display] on the scope, wait for the #Avgs value (top right corner of screen)
to return to 256, and then record the DMM voltage reading as V
DMM+
and the scope
Vavg reading as V
Scope+
in the Offset Gain Test section of the Performance Test Record.