Chapter 3: Testing Performance
DC Gain Accuracy Test
25
b When the Acquisition Setup window appears, enable averaging and set the # of
averages to 256 as shown below.
3 Set the Cal Out voltage (V
Cal Out
) to +30 mV as follows:
a Select Utilities > Calibration Output.
b Change the Signal Output function to DC (top left corner).
c Set the Level to 30 mV.
d Click Close.
4 Set the channel 1 vertical sensitivity value to 10 mV/div. This can be done either using
the front panel control or:
a Pull down the Setup menu and select Channel 1 or click the Channel 1 setup icon.
b Change the vertical sensitivity of channel 1 to 10 mV/div.
Vista de pagina 24