Agilent Technologies Infiniium 90000 Q-Series Especificaciones Pagina 30

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 136
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 29
Chapter 3: Testing Performance
Analog Bandwidth -- Maximum Frequency Test
30
Connections for all oscilloscope models, channels 1- 4
Notes
Connect output 1 of the 11667C splitter to the scope Channel n input directly using the 11901A adapter
and a connector saver, without any additional cabling or adapters.
Connect the power sensor directly to output 2 of the power splitter using the 11901C adapter without
any additional cabling or adapters.
Connect the microwave cable directly to output 3 of the power splitter using the 11901C adapter
without any additional cabling or adapters.
Minimize the use of other adapters.
Ensure that 2.4 mm and 3.5 mm connectors are tightened properly:
8 in-lbs (90 N-cm) for 3.5 mm
8 in-lbs (90 N-cm) for 2.4 mm
Power sensor
cable
Power sensor
N1922A
Power splitter
11667C
11901C adapter
11901A adapter
Microwave cable
Microwave CW
Generator E8257D
Power meter
N1912A
Connection saver
Oscilloscope
Vista de pagina 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 135 136

Comentarios a estos manuales

Sin comentarios