Chapter 4: Troubleshooting
Power Supply Troubleshooting
54
The second set of LEDs is located to the left of the first set and slightly lower, below
the J12002 connector (see Figure 4- 6).
Figure 4-6 Second set of LEDs
The following table shows each LED in this second set, the type of fault it indicates,
and what to do if that LED indicator is illuminated. When identifying which acquisition
assembly to replace, remember that the acquisition assembly nearest the bottom of the
oscilloscope is for channel 1. The one above it is for channel 2. The next higher is for
channel 3, and the top acquisition assembly is for channel 4.
PS71 PS72 PS73 PS74
ON ON ON ON Supply late
Replace Backplane Assembly
ON ON ON OFF Supply late
ON ON OFF ON Supply late
ON ON OFF OFF Supply late
ON OFF ON ON Under-voltage
ON OFF ON OFF Under-voltage
ON OFF OFF ON Under-voltage
ON OFF OFF OFF Under-voltage
OFF OFF ON ON Over-voltage
OFF OFF ON OFF Over-voltage
OFF OFF OFF ON Over-voltage
OFF OFF OFF OFF Over-voltage
OFF ON X X Power button pushed at wrong time Unplug scope, wait 30 seconds,
plug in, power up
NOTE
There is no correlation between the short name of each LED and the type of
fault. Use the table to determine the fault and how to address the problem.
LED Type of Fault What To Do