Chapter 5: Replacing Assemblies
Removing and Replacing the Front Panel Assembly and Bottom Panel
82
9
Disconnect the Aux Out and Cal Out cables from both the front panel and the
backplane board. Pull straight back to remove the cables from the backplane
board. CAUTION: Be sure to pull straight back to avoid bending or breaking
the connection to the backplane board. When reconnecting, gently reinsert each
cable. You will hear a snap when it is fully seated.
Figure 5-25
Figure 5-26
The front panel is now disconnected.
Removing a RealEdge Assembly
If you are working on a 50 or 62 GHz model, it will have a RealEdge assembly.
The following steps show how to remove it. Be sure to remove the RealEdge
assembly before removing the front panel to avoid damaging the assembly.
1 Disconnect the two input cables from the RealEdge modules and from the
RealEdge modules to the RealEdge auxiliary module.
Figure 5-27