Agilent Technologies Infiniium 90000 Q-Series Especificaciones Pagina 81

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 136
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 80
Chapter 5: Replacing Assemblies
Removing and Replacing the Front Panel Assembly and Bottom Panel
81
6
Disconnect each input cable from each of the four attenuator connectors using
a 5/16- inch Hex torque wrench. When reconnecting, torque to 8 in- lbs. The
length of the cable will guide the installation (longest cable on left, shortest
cable on right).
Figure 5-22
7
Disconnect the attenuator power cable from each input channel attenuator.
Figure 5-23
8
Disconnect the Probe Comp cable from the connector on the backplane board.
Figure 5-24