Chapter 3: Testing Performance
Performance Test Record
39
Performance Test Record
Offset Accuracy Test
Zero Error Test
Agilent Technologies
Agilent 90000 Q-Series Oscilloscopes
Model Number _____________________ Tested by___________________
Serial Number ___________________________ Work Order No.___________________
Recommended Test Interval - 1 Year/2000 hours Date___________________
Recommended next test date ___________________ Ambient temperature ___________________
Vertical
Sensitivity
Test Limits Channel 1 Channel 2 Channel 3 Channel 4 Channel 1R Channel 3R
10 mV/div -1.8 mV to +1.8 mV
20 mV/div -2.6 mV to +2.6 mV
50 mV/div -5.0 mV to +5.0 mV
100 mV/div -9.0 mV to +9.0 mV
200 mV/div -17.0 mV to +17.0 mV
500 mV/div -41.0 mV to +41.0 mV
1 V/div -81.0 mV to +81.0 mV