© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 24
Installing DUT Power Supplies
Installing the DUT Power Supply in a 317X
System
These procedures describe how to install the DUT
power supply in a 317X testhead. If you are installing an
6634A/B DUT power supply, there are some additional
steps described below to install a radiation shield
(E4034-00601), which is integrated into a new power
supply support rail.
1 Using a #2 Pozidriv screwdriver, remove the side
handles from both sides of the power supply. Discard
these parts.
2 Remove the adhesive-backed trim strips from the
sides of the power supply’s front frame. Then, using
a #2 Pozidriv screwdriver, install rack-mount ears on
both sides of the power supply. Pan-head screws are
included with the rack-mount ears.
3 On the top of the testhead is a sheet metal cover
(E1170-01208) over the bank containing the power
supplies. Remove the four T15 screws that attach the
cover to the testhead frame.
4 Using a 7/16-inch box or open-end wrench, remove
the hex-head screws, split lockwashers, and flat
washers from the rear of the testhead (see Figure 8
on page 25).
5 Using a #2 Pozidriv offset or ratchet screwdriver,
remove the DUT power supply mounting bars from
the support rails shown in Figure 8. Save the
flat-head Pozidriv screws.