© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 30
Installing DUT Power Supplies
Installing the DUT Power Supply in a 327X
System
These procedures describe how to install the DUT
power supply in a 327X testhead. If you are installing a
Faceless 6624A or an 6634A/B DUT power supply,
there are some additional procedures, as indicated
below. If it is not already present, a radiation shield
(E3770-00601) must be mounted onto the existing
support rails. Without this shield in place, the power
supply might radiate into nearby ASRU cards causing
measurement failures. This procedure is to be done only
by Agilent 3070 / 79000 service-trained customer
engineers.
1 Use a #2 Pozidriv screwdriver to remove the side
handles from both sides of the power supply.
Discard these parts.
2 Determine the location for the power supply using
Figure 11 on page 31.