Chapter 2 Service
The Agilent 53131A/132A Calibration Procedures
2-20 Assembly-Level Service Guide
2
CAL: TI QUIK? Calibration
1 Press any one of the arrow keys until CAL: TI QUIK? is displayed.
2 Press Enter key.
A scrolling message with instructions appears. Follow the instruction by performing
the following procedure
3 Connect the Agilent 8130A Pulse Generator output to Channel 1 of the Counter
as shown in Figure 2-3.
Equipment
Agilent 8130A Pulse Generator (or equivalent)
Agilent 10503A BNC Cable
Figure 2-3. Calibration Setup for CAL: TI QUIK?
4 Set the Agilent 8130A as follows:
PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 V Low : 0.50 V
Input Mode: Normal
NOTE Any external timebase connected to the Counter is ignored during this calibration.
HP53131A/132A
Counter
Channel 1
HP 8130A
Pulse Generator
Output