Chapter 1 Performance Tests
Agilent 53131A/132A Performance Test Record (Page 4 of 6)
1-44 Assembly-Level Service Guide
1
Agilent 53131A/132A Performance Test Record
(Page 4 of 6)
Complete Performance Tests (Continued)
Test
Number Test Description Minimum
Actual Reading
Maximum
4 200 MHz Test
(Agilent 53131A):
Channel 1
Channel 2
200 MHz Test
(Agilent 53132A):
Channel 1
Channel 2
199.999,999,87 MHz
199.999,999,87 MHz
199.999,999,970 MHz
199.999,999,970 MHz
9. __________
10. __________
11. __________
12. __________
200.000,000,13 MHz
200.000,000,13 MHz
200.000,000,03 MHz
200.000,000,03 MHz
4 225 MHz Test
(Agilent 53131A):
Channel 1
Channel 2
225 MHz Test
(Agilent 53132A):
Channel 1
Channel 2
224.999,999,85 MHz
224.999,999,85 MHz
224.999,999,965 MHz
224.999,999,965 MHz
13. __________
14. __________
15. __________
16. __________
225.000,000,15 MHz
225.000,000,15 MHz
225.000,000,035 MHz
225.000,000,035 MHz
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (Agilent 53131A/132A Only)
Actual Reading Specification
100 MHz–2.5 GHz
2.7–3.0 GHz
200 MHz–5.0 GHz
200 MHz–12.4 GHz
1. __________
2. __________
3. __________
4. __________
27 dBm 10 mV rms
21 dBm 20 mV rms
23 dBm 16 mV rms
23 dBm 16 mV rms