Chapter 1 Performance Tests
Agilent 53131A/132A Performance Test Record (Page 5 of 6)
Assembly-Level Service Guide 1-45
1
Agilent 53131A/132A Performance Test Record
(Page 5 of 6)
Complete Performance Tests (Continued)
Test
Number Test Description Minimum
Actual Reading
Maximum
6 Option 030 Channel 3
Frequency Accuracy
(Agilent 53131A/132A
Only)
100 MHz Test
(Agilent 53131A):
Channel 3
100 MHz Test
(Agilent 53132A):
Channel 3
99.999,999,93 MHz
99.999, 999,980 MHz
1. __________
2. __________
100.000,000,07 MHz
100.000,000,020 MHz
6 3.0 GHz Test
(Agilent 53131A):
Channel 3
3.0 GHz Test
(Agilent 53132A):
Channel 3
2,999.999,998 MHz
2,999.999,999,60 MHz
3. __________
4. __________
3,000.000,002 MHz
3,000.000, 000,4 MHz
6 5.0 GHz Test
(Agilent 53131A):
Channel 3
5.0 GHz Test
(Agilent 53132A):
Channel 3
4,999.999,997,0 MHz
4,999.999,999,3 MHz
5. __________
6. __________
5,000.000,003,0 MHz
5,000.000, 000,7 MHz