Agilent Technologies HP 3325B Manual de instrucciones Pagina 57

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 252
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 56
Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (Agilent
53131A/132A Only)
Assembly-Level Service Guide 1-33
1
200 MHz to 5.0 GHz Sensitivity for Agilent 53131A/132A (Option
050 only)
1 Connect the 8340B Sweep Oscillator to Channel 3 of the counter.
2 Change the settings of the Sweep Oscillator to 5.0 GHz, 40 dBm.
3 Increase the power level in 1 dB steps until the Counter displays a stable reading
of approximately 5.0 GHz.
The Counter should display 5.0 GHz at 23 dBm.
NOTE Use a power meter and sensor to ensure the accuracy of the power being input.
4 Sweep the frequencies from 200 MHz to 5.0 GHz.
The Counter should read frequencies from 200 MHz to 5.0 GHz at
≤−23 dBm.
5 Record the actual reading in the Performance Test Record (Test 5, Line 3).
6 Disconnect the test setup.
Vista de pagina 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 251 252

Comentarios a estos manuales

Sin comentarios