Chapter 6 Backdating
The Agilent 53131A Calibration Procedures
Assembly-Level Service Guide 6-13
6
CHANGE 1 (Firmware Revisions 3317, 3335, 3402, 3427) (Continued)
Equipment
Agilent 8130A Pulse Generator (or equivalent)
Agilent 10503A BNC Cable
Figure 2-3. TI Calibration Setup
4 Set the Agilent 8130A as follows:
PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 VLow : 0.50 V
Input Mode: Normal
NOTE Any external timebase connected to the Counter is ignored during this calibration.
5 Press Enter key.
The Counter displays CALIBRATING, and then it should display TI CAL PASS.
If the fail message is displayed, refer to the troubleshooting section in this chapter.
CHANGE 1 (Firmware Revisions 3317, 3335, 3402, 3427) (Continued)
6 Disconnect the calibration setup.
This completes the TI QUIK? calibration procedures.
HP53131A
Counter
Channel 1
HP 8130A
Pulse Generator
Output