Chapter 2 Service
About the Agilent 53181A Calibration Menu
Assembly-Level Service Guide 2-27
2
The Agilent 53181A Calibration Menu Tree
NOTE Turn power off, press and hold Scale & Offset key, then press POWER key to access
this menu.
3
This menu item appears and calibration is permitted only if calibration is unsecured. Enter in the correct code
to change calibration to secure.
4
Timebase can be automatically calibrated only if the timebase option is installed.
Scale &
Offset
On / Stby
POWER
CAL SECURE CAL UNSECURE
3
CAL: OFFS1?
CAL: GAIN1?
CAL: TIMEBAS?
CODE: 0
CAL COUNT?
HELP: CAL?
HELP: SECURE?
HELP: CODE?
4