Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (Agilent 53181A Only)
1-54 Assembly-Level Service Guide
1
100 MHz to 200 MHz Sensitivity for Agilent 53181A
1 Change the signal generator settings to 110 MHz at 30 dBm.
2 Increase the power level by 0.1 dB steps until the Counter displays a stable 110
MHz reading.
The Counter should read 200 MHz when the input signal level
is 17.5 dBm (30 mVrms).
3 Sweep the signal generator frequency from 100 MHz to 200 MHz.
The Counter should read frequencies from 100 MHz to 200 MHz when the input
signal level is 17.5 dBm (30 mVrms).
4 Record the actual reading in the Performance Test Record (Test 2, Line 2).
200 MHz to 225 MHz Sensitivity for Agilent 53181A
1 Change the signal generator settings to 200 MHz at 30 dBm.
2 Increase the power level by 0.1 dB steps until the Counter displays a stable 200
MHz reading.
The Counter should read 200 MHz when the input signal level is
15.0 dBm (40 mVrms).
3 Sweep the signal generator frequency from 200 MHz to 225 MHz.
The Counter should read frequencies from 200 MHz to 225 MHz when the input
signal level is 15.0 dBm (40 mVrms).
4 Record the actual reading in the Performance Test Record (Test 2, Line 3).
5 Disconnect the test setup.