Chapter 1 Performance Tests
Agilent 53131A/132A Performance Test Record (Page 2 of 6)
1-42 Assembly-Level Service Guide
1
Agilent 53131A/132A Performance Test Record
(Page 2 of 6)
Complete Performance Tests
Test Number
Test Description Minimum Actual Reading Maximum
1 Time Interval
(Agilent 53131A/132A
Only):
For Agilent 53131A
For Agilent 53132A
0.1460 µs
0.1475 µs
1. __________
2. __________
0.1540 µs
0.1525 µs
2 Trigger Level:
Channel 1
Channel 2
(Agilent 53131A/132A Only)
15 mV
15 mV
1. __________
2. __________
+15 mV
+15 mV
Test 3: Channels 1 and 2 Frequency Sensitivity (Agilent 53131A/132A Only)
Standard (Front Panel Terminals) Actual Reading Specification
100 kHz–100 MHz, Channel 1
100 kHz–100 MHz, Channel 2
1. __________
2. __________
21 dBm (20 mVrms)
21 dBm (20 mVrms)
100 MHz–200 MHz, Channel 2
100 MHz–200 MHz, Channel 1
3. __________
4. __________
17.5 dBm (30 mVrms)
17.5 dBm (30 mVrms)
200 MHz–225 MHz, Channel 1
200 MHz–225 MHz, Channel 2
5. __________
6. __________
15.0 dBm (40 mVrms)
15.0 dBm (40 mVrms)
Option 060 Rear Terminals Actual Reading Specification
100 kHz–225 MHz, Channel 1
100 kHz–225 MHz, Channel 2
1. __________
2. __________
9.5 dBm (75 mVrms)
9.5 dBm (75 mVrms)