SCPI Commands Reference Common Commands
22 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
Low Level :VOLT[0|1|2]:LOW
<value>
-50 mV -2.00 V … +2.95 V
High :VOLT[0|1|2]:HIGH
<value>
50 mV -1.95 V … +3.00 V
Offset :VOLT[0|1|2]:OFFS
<value>
0 mV -1.975 V … +2.975 V
Amplitude :VOLT[0|1|2] <value> 100 mV 50 mV … 2.00 V
Term. Voltage :VOLT[0|1|2]:TERM
<value>
0 mV -2.00 V … +3.00 V
Limit to current levels :VOLT[1|2]:LIM <value> OFF OFF, ON
Data Length :DIG:PATT:LENG <value> 32 32 … 8192 (in increments of 32)
Clock Input
Termination
:TRIG:TERM:STATE
<value>
OFF ON, OFF
Clock Input Term.
Voltage
:TRIG:TERM <value> 0 mV -2.00 … +3.00 V
Trigger Output Mode :OUTP0:SOUR <value> PERiodic PERiodic, BITStream
Trigger Output Divider :OUTP0:DIV <value> 1 1, 2, 3, … (2
31
- 1)
Trigger Output High see High Level 50 mV -1.95 V … +3.00 V
Trigger Output Low see Low Level -50 mV -2.00 … +2.95 V
Trigger Output Term.
Voltage
see Term. Voltage 0 mV -2.00 … +3.00 V
Start Input Start Mode :ARM:SOUR <value> IMMediate IMMediate|MANual|EXTernal
Start Input Term.
Voltage
:ARM:TERM <value> 0 mV -2.00 … +3.00 V
Start Input Threshold :ARM:LEV <value> 100 mV -2.00 … +3.00 V
Start Input Start On :ARM:SLOP <value> POS (rising) POS, NEG (rising/falling)
Parameter SCPI Command Reset Value Range