Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 73
:STATus:QUEStionable:VOLTage
This command tree accesses the QUEStionable:VOLTage status
group.
The QUEStionable:VOLTage status group monitors the currently
programmed output voltage levels against their specified ranges.
The following commands are used to access the registers within
the status group.
:STATus:QUEStionable:VOLTage[:EVENt]?
Syntax
:STAT:QUES:VOLT[:EVEN]?
Form Query
Description Reads the event register in the QUEStionable:VOLTage status
group.
Parameter
*RST value
:STATus:QUEStionable:VOLTage:CONDition?
Syntax
:STAT:QUES:VOLT:COND?
Form Query
Description Reads the condition register in the QUEStionable:VOLTage status
group.
Parameter
*RST value
:STATus:QUEStionable:VOLTage:ENABle
Syntax
:STAT:QUES:VOLT:ENAB[?]
Form Set & Query
Description Sets or queries the enable register in the QUEStionable:VOLTage
status group.
Parameter Numeric
*RST value