Agilent Technologies 81134A Manual de servicio Pagina 25

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 104
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 24
DIGital Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 25
DIGital Commands
The following table shows the Agilent 81133A/81134A Pulse
Generator DIGital commands:
Command Parameter Description
:DIGital[1|2]
[:STIMulus]
:PATTern
[:DATa][?] <data>, [HEX |
BIN | DUAL]
Sets/reads data in hexadecimal
(default), binary or dual format;
this command is for data pat
-
terns with maximum 8192 bits
:LDATa <data>
Sets data in hexadecimal for-
mat; this command is for data
patterns larger than 8192 bits
:LENGth[?]
Numeric Sets/reads data pattern length
in bits (32 ... 8192) in steps of 32
:SIGNal
:FORMat[?] RZ | NRZ | R1
Sets/reads the signal mode
:POLarity[?] NORMal |
COMPlement |
INVerted
Sets/reads data polarity
:CROSsover
:[VALUE][?]
Numeric Sets/reads crossover
:STATe[?] ON | OFF | 1 |
0
Switches crossover on/off
Vista de pagina 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 103 104

Comentarios a estos manuales

Sin comentarios