SCPI Commands Reference SYSTem Commands
84 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
:SYSTem:COMMunicate:LAN[:SELF]
:ADDRess
Syntax
:SYST:COMM:LAN[:SELF]:ADDR[?]
Form Set & Query
Description Sets the instrument's IP address.
NOTE This parameter must only be set if DHCP is not available. See
“:SYSTem:COMMunicate:LAN[:SELF]:DHCP” on page 83.
Parameter String <no>.<no>.<no>.<no> in quotes, where <no> is in the range 1
… 255.
*RST value
Example
:SYST:COMM:LAN:ADDR "150.215.17.9"
:SYSTem:COMMunicate:LAN[:SELF]
:SMASk
Syntax
:SYST:COMM:LAN[:SELF]:SMAS[?]
Form Set & Query
Description Sets the instrument’s subnet mask.
NOTE This parameter must only be set if DHCP is not available. See
“:SYSTem:COMMunicate:LAN[:SELF]:DHCP” on page 83.
Parameter String <no>.<no>.<no>.<no> in quotes, where <no> is in the range 0
… 255.
*RST value
Example
:SYST:COMM:LAN:SMAS "255.255.240.000"