Troubleshooting
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 99
If the trigger amplitude is increased, or
if the offset is decreased.
• "low level of channel [1|2] is below minimum"
May happen:
If the channel amplitude is increased, or
if the offset is decreased.
• "high level of trigger exceeds maximum"
May happen:
If the trigger amplitude is increased, or
if the offset is increased.
• "high level of channel [1|2] exceeds maximum"
May happen:
If the channel amplitude is increased, or
if the offset is increased.
• "high level of trigger is lower than low level"
May happen:
If the trigger amplitude is set to a negative value, or
if high level and low level are set in parallel by list of
semicolon- separated SCPI commands.
• "high level of channel [1|2] is lower than low level"
May happen:
If the channel amplitude is set to a negative value, or
if high level and low levels are set in parallel by list of
semicolon- separated SCPI commands.
Instrument not Operable via LAN
If the instrument cannot be programmed via LAN, make sure that
you have installed the latest Agilent I/O library on your computer.
Visit the Agilent web site for the newest version.