SCPI Commands Reference SOURce Commands
60 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
[:SOURce]:VOLTage[0|1|2][:LEVel]
[:IMMediate][:AMPLitude]
Syntax
[:SOUR]:VOLT[0|1|2][:LEV][:IMM][:AMPL][?]
Form Set & Query
Description Programs the amplitude of the output signal for the trigger output
and the channels.
Parameter Numeric [uV|mV|V] |MIN|MAX
Value coupling High = Offset + (Amplitude / 2)
Low = Offset – (Amplitude / 2)
Range coupling Offset
*RST value Trigger output (0): 100 mV
Channels 1 and 2: 100 mV
Specified limits Trigger output (0): 50 mV... 2.0 V
Channels 1 and 2: 50 mV… 2.0 V
Absolute limits Trigger output (0): 2.0 V
Channels 1 and 2: 2.0 V
Example Set Trigger Output amplitude to 1 V.
:VOLT0 1V