Agilent Technologies 81134A Manual de servicio Pagina 81

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 104
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 80
SYSTem Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 81
:SYSTem:ERRor?
Syntax
:SYST:ERR?
Form Query
Description This command is used to read the Agilent 81133A/81134A Pulse
Generator error queue. The Agilent 81133A/81134A Pulse
Generator error queue can store up to 32 error codes on a first-
in- first- out basis. When you read the error queue, the error
number and associated message are put into the instrument’s
output buffer.
If the error queue is empty, the value 0 is returned, meaning No
Error. If the queue overflows at any time, the last error code is
discarded and replaced with –350 meaning Queue Overflow.
Refer to “Troubleshooting” on page 97 for a list of the error
messages.
Parameter
*RST value
:SYSTem:PRESet
Syntax
:SYST:PRES
Form Event
Description This command is equivalent to
*RST
, except that there is no change
to the
:DISPlay[:WINDow][:STATe]
. Use this command instead of
*RST
if you want the display to remain switched off during program
execution.
Parameter
*RST value
Vista de pagina 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 103 104

Comentarios a estos manuales

Sin comentarios