SCPI Commands Reference SOURce Commands
66 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
[:SOURce]:VOLTage[1|2]:LIMit:STATe
Syntax
[:SOUR]:VOLT[1|2]:LIM:STAT[?]
Form Set & Query
Description Switches the “Limit to current values” output mode on or off.
When you switch on Limited output mode the current high-level
and low- level parameters are taken as limit values restricting the
available ranges of all output- level parameters. You cannot program
the output- levels beyond these temporary limits, until you switch
off Limited output mode. The limits apply whether you program
high/low levels or amplitude/offset levels.
Parameter ON|OFF|1|0
*RST value OFF
Example Switch on Channel 1 Limited output mode.
:VOLT1:LIM:STAT ON