SOURce Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 53
[:SOURce]:PHASe[:ADJ][1|2]
Syntax
[:SOUR]:PHAS[:ADJ][1|2][?]
Form Set & Query
Description This command programs the pulse phase for a channel.
Parameter Numeric | MIN|MAX
Parameter Suffix DEG or RAD. A parameter without suffix is interpreted as degrees.
Functional coupling Programming the pulse phase also executes
[:SOURce][:PULSe]:HOLD PHASe
so that the pulse phase is held
constant when the signal frequency is changed.
Value coupling Delay = (Phase / 360) * Period
Range coupling Deskew
*RST value 0.0
Specified limits –6000° … +279000°, constrained by delay and period limits.
Absolute limits –6000° … +279000°, constrained by delay and period limits.
Example Set channel 1 phase delay to –180°.
:PHAS1 -180