Agilent Technologies B1500A Guía de usuario Pagina 155

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 628
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 154
Agilent B1500 Users Guide, Edition 7 3- 25
Installation
Installing Accessories
To Connect GNDU Adapter
The ground unit (GNDU) output is the single triaxial connector. Agilent
N1254A-100 GNDU to Kelvin Adapter enables to change it to the dual triaxial
connector. Connect the adapter to the GNDU connector on the rear panel, then
connect a Kelvin triaxial cable to extend the GNDU terminal. Maximum current is
restricted when the Kelvin triaxial cable is used. See Table 3-4.
Figure 3-6 Agilent N1254A-100 GNDU to Kelvin Adapter
Table 3-4 Kelvin Triaxial Cable and Maximum Current
Model No. or
Op
tion It
em
Description
Maximum
Cu
rrent
16493K-001 1.5 m Kelvin Triaxial Cable 3.0 A
16493K-002 3 m Kelvin Triaxial Cable 2.6 A
'XDO7ULD[LDOIHPDOH
7R*1'8FRQQHFWRU 7R..HOYLQWULD[LDOFDEOH
$JLOHQW1$*1'8WR.HOYLQ$GDSWHU
7ULD[LDOPDOH
&RPPRQ
6HQVH
)RUFH
Vista de pagina 154
1 2 ... 150 151 152 153 154 155 156 157 158 159 160 ... 627 628

Comentarios a estos manuales

Sin comentarios