Vista de pagina 569
10-18 Agilent B1500 User’s Guide, Edition 7
Application Library and Utilities
QSCV Measurement Supplemental Data
Table 10-2 Constant Value for Calculating the Measurement Accuracy: HRSMU/ASU
Measurement
Range Constant
Voltage Output Range
QSCV Operating
Mode
0.5 V 2 V 5 V 20 V 40 V 100 V
10 pA Ap 0.47 0.47 0.47 0.47 0.48 0.48
Bp 0.014 0.029 0.051 0.15 0.30 0.72 Normal
0.009 0.011 0.015 0.03 0.06 0.13 4155C/4156C
compatible
Cp 0.013
Dp 0.001
Ao 0.04 Normal
0.07 4155C/4156C
compatible
Bo 1.3E-16
Boc 1.1E-05
Co 5.4E-15
Coc 2.4E-05
Do 1.3E-16
Doc 1.1E-05
100 pA Ap 0.31 0.31 0.31 0.31 0.32 0.32
Bp 0.014 0.029 0.051 0.153 0.298 0.720 Normal
0.009 0.011 0.015 0.033 0.056 0.131 4155C/4156C
compatible
Cp 0.009
Dp 0.002
Ao 0.03 Normal
0.071 4155C/4156C
compatible
Bo 4.7E-16
Boc 3.5E-05
Co 1.3E-14
Coc 9.8E-05
Do 4.7E-16
Doc 3.5E-05