Agilent B1500 Users Guide, Edition 7 2- 25
Introduction
Measurement Units
Table 2-9 HRSMU Output Voltage Ranges and Resolutions
Table 2-10 HRSMU Output Current Ranges and
Resolutions
Range Output Value
Setting
Resolution
Ma
ximum
Current
0.5 V 0 |V| 0.5 V 25 μV ± 100 mA
2 V 0 |V| 2 V 100 μV ± 100 mA
5 V 0 |V| 5 V 250 μV ± 100 mA
20 V 0 |V| 20 V 1 mV ± 100 mA
40 V 0 |V| 20 V 2 mV ± 100 mA
20 V < |V| 40 V 2 mV ± 50 mA
100 V 0 |V| 20 V 5 mV ± 100 mA
20 V < |V| 40 V 5 mV ± 50 mA
40 V < |V| 100 V 5 mV ± 20 mA
Range Output Value
Setting
Res
olution
Maximum
Voltage
1 pA
a
a. Available when the E5288A Atto Sense and Switch Unit (ASU) is used.
0 |I| 1.15 pA 1 fA ± 100 V
10 pA 0 |I| 11.5 pA 5 fA ± 100 V
100 pA 0 |I| 115 pA 5 fA ± 100 V
1 nA 0 |I| 1.15 nA 50 fA ± 100 V
10 nA 0 |I| 11.5 nA 500 fA ± 100 V
100 nA 0 |I| 115 nA 5 pA ± 100 V
1 μA 0 |I| 1.15 μA 50 pA ± 100 V
10 μA 0 |I| 11.5 μA 500 pA ± 100 V
100 μA 0 |I| 115 μA
5 nA ± 100 V
1 mA 0 |I| 1.15 mA 50 nA ± 100 V
10 mA 0 |I| 11.5 mA 500 nA ± 100 V
100 mA 0 |I| 20 mA 5 μA ± 100 V
20 mA < |I| 50 mA 5 μA ± 40 V
50 mA < |I| 100 mA 5 μA ± 20 V
Vista de pagina 72