Agilent B1500 Users Guide, Edition 7 2- 21
Introduction
Measurement Units
B1511A Medium Power SMU
This section describes typical specification of the medium power source/monitor
unit (MPSMU) for Agilent B1500A.
Maximum voltage, current, output power: ± 100 V, ± 100 mA, 2 W
Minimum range: 0.5 V, 1 nA
Output/measurement value and resolution: see Table 2-5 to Table 2-8.
Figure 2-4 MPSMU Output and Measurement Ranges











9ROWDJH9
&XUUHQWP$
