7- 56 Agilent B1500 User’s Guide, Edition 7
Function Details
SMU Pulse
SMU Pulse
The source/monitor unit (SMU) can apply voltage or current pulse. The pulse width
and pulse period must be the following values.
Pulse width 0.5 ms to 2 s, 0.1 ms resolution
Pulse period 5 ms to 5 s, 0.1 ms resolution
The setup values must satisfy the following conditions.
period w
idth + 2 ms (f
or pulse width 100 ms)
period width + 10 ms
(for pulse width > 100 ms)
To Set Pulse Output
The following points must be noted when setting the pulse output.
Cannot use the 1 nA output range and lower ranges.
Cannot use multiple measurement channels.
The measurement channel uses the co
mpliance range. The chann
el cannot
perform the auto ranging operation or the limited auto ranging operation.