1-10 Agilent B1500 User’s Guide, Edition 7
Getting Started
To Use Application Test Mode
To Use Application Test Mode
For the screen as shown in Figure 1-4, perform the following procedure.
Figure 1-4 Display Example of EasyEXPERT Main Screen
1. Select a test definition on the EasyEXPERT main screen while the Application
Test tab has been selected. The tab is in the leftmost column on the screen.
a. Set the Category area to change the test definitions listed in the Library area.
For example, check the CMOS only.
b. Select a test definition at the Library area. For example, click Id-Vg and
Select button sequentially. The Id-Vg screen is displayed. See Figure 1-5.
2. Change the test conditions at the Test Parameters area on the main screen as
shown in Figure 1-5.
For example, the following steps change the sweep range of voltage output.
a. Set VgStart to 0 V and VgStep to 0.2 V.
b. VsubsStep to -1.0 V.
Vista de pagina 41