Agilent Technologies B1500A Guía de usuario Pagina 88

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 628
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 87
2- 40 Agilent B1500 User’s Guide, Edition 7
Introduction
Specifications
ASU
ASU: Atto Sense and Switch Unit
Maximum voltage, AUX path
100 V (AUX input to AUX common)
100 V (AUX input to circuit common)
42 V (AUX common to circuit common)
Maximum current, AUX path
0.5 A (AUX input to Force output)
Band width (at -3 dB), AUX port
30 MHz, supplemental information
NOTE Connect ASU to dedicated HRSMU
The specifications are satisfied and guaranteed for the exclusive combination of the
ASU and the HRSMU. So confirm the serial number of the ASU and connect it to
the dedicated HRSMU properly.
NOTE Connection path control
Disable source output of the SMU and other instrument connected to the ASU
inputs before controlling the ASU connection path.
Vista de pagina 87