Agilent Technologies B1500A Guía de usuario Pagina 593

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 628
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 592
Agilent B1500 Users Guide, Edition 7 10-41
Application Library and Utilities
QSCV Measurement Supplemental Data
Figure 10-23 Calculation Example of Measurement Accuracy: HRSMU/ASU
Conditions:
QSCV Operating Mode: Normal
Measurement Range: 1 μA
Output Range: 20 V
Integration Time: 2, 1, 0.5, 0.3 s corresponding to the lines from top to bottom
Equivalent Parallel Resistance of DUT: 100 M ohm
Guard Capacitance of Measurement Path: 200 pF
0
2
4
6
8
10
12
14
0.001 0.01 0.1 1 10
QSCVMeasurement Voltage (V)
Reading Accuracy (%)
1.E-14
1.E-13
1.E-12
1.E-11
1.E-10
1.E-09
0.001 0.01 0.1 1 10
QSCVMeasurement Voltage (V)
Offset Accuracy (F)
Vista de pagina 592
1 2 ... 588 589 590 591 592 593 594 595 596 597 598 ... 627 628

Comentarios a estos manuales

Sin comentarios