Root (:) Commands 6
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 157
:AUToscale:FDEBug
(see page 790)
Command Syntax
:AUToscale:FDEBug <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :AUToscale:FDEBug command turns fast debug auto scaling on or off.
The Fast Debug option changes the behavior of :AUToscale to let you make
quick visual comparisons to determine whether the signal being probed is
a DC voltage, ground, or an active AC signal.
Channel coupling is maintained for easy viewing of oscillating signals.
Query Syntax
:AUToscale:FDEBug?
The :AUToscale:FDEBug? query returns the current autoscale fast debug
setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to Root (:) Commands" on page 150