Agilent Technologies 2000 X-Series Manual de servicio Pagina 376

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 930
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 375
376 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
20 :MEASure Commands
:MEASure:WINDow
(see page 790)
Command Syntax
:MEASure:WINDow <type>
<type> ::= {MAIN | ZOOM | AUTO}
When the zoomed time base is displayed, the :MEASure:WINDow command
lets you specify the measurement window:
MAIN — the measurement window is the upper, Main window.
ZOOM — the measurement window is the lower, Zoom window.
AUTO — the measurement is attempted in the lower, Zoom window; if it
cannot be made there, the upper, Main window is used.
Query Syntax
:MEASure:WINDow?
The :MEASure:WINDow? query returns the current measurement window
setting.
Return Format
<type><NL>
<type> ::= {MAIN | ZOOM | AUTO}
See Also "Introduction to :MEASure Commands" on page 340
":MEASure:SOURce" on page 361
Vista de pagina 375
1 2 ... 371 372 373 374 375 376 377 378 379 380 381 ... 929 930

Comentarios a estos manuales

Sin comentarios