Agilent Technologies 2000 X-Series Manual de servicio Pagina 264

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 930
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 263
264 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
14 :DVM Commands
:DVM:ARANge
(see page 790)
Command Syntax
:DVM:ARANge <setting>
<setting> ::= {{OFF | 0} | {ON | 1}}
If the selected digital voltmeter (DVM) source channel is not used in
oscilloscope triggering, the :DVM:ARANge command turns the digital
voltmeter's Auto Range capability on or off.
When on, the DVM channel's vertical scale, vertical (ground level)
position, and trigger (threshold voltage) level (used for the counter
frequency measurement) are automatically adjusted.
The Auto Range capability overrides attempted adjustments of the
channel's vertical scale and position.
When off, you can adjust the channel's vertical scale and position
normally.
Query Syntax
:DVM:ARANge?
The :DVM:ARANge? query returns a flag indicating whether the digital
voltmeter's Auto Range capability is on or off.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also ":DVM:SOURce" on page 269
":DVM:ENABle" on page 266
":DVM:MODE" on page 268