400 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
21 :MTESt Commands
:MTESt:RMODe:FACTion:STOP
(see page 790)
Command Syntax
:MTESt:RMODe:FACTion:STOP <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:STOP command sets stopping on a mask
failure on or off. When this setting is ON and a mask violation is detected,
the mask test is stopped and the acquisition system is stopped.
Query Syntax
:MTESt:RMODe:FACTion:STOP?
The :MTESt:RMODe:FACTion:STOP? query returns the current mask failure
stop setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also "Introduction to :MTESt Commands" on page 379
":MTESt:RMODe:FACTion:MEASure" on page 397
":MTESt:RMODe:FACTion:PRINt" on page 398
":MTESt:RMODe:FACTion:SAVE" on page 399