Agilent Technologies 2000 X-Series Manual de servicio Pagina 583

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 930
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 582
:TIMebase Commands 28
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 583
:TIMebase:SCALe
(see page 790)
Command Syntax
:TIMebase:SCALe <scale_value>
<scale_value> ::= time/div in seconds in NR3 format
The :TIMebase:SCALe command sets the horizontal scale or units per
division for the main window.
Query Syntax
:TIMebase:SCALe?
The :TIMebase:SCALe? query returns the current horizontal scale setting in
seconds per division for the main window.
Return Format
<scale_value><NL>
<scale_value> ::= time/div in seconds in NR3 format
See Also "Introduction to :TIMebase Commands" on page 578
":TIMebase:RANGe" on page 581
":TIMebase:WINDow:SCALe" on page 587
":TIMebase:WINDow:RANGe" on page 586
Vista de pagina 582
1 2 ... 578 579 580 581 582 583 584 585 586 587 588 ... 929 930