484 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
:SBUS<n>:LIN:SOURce
(see page 790)
Command Syntax
:SBUS<n>:LIN:SOURce <source>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital<d>} for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :SBUS<n>:LIN:SOURce command sets the source for the LIN signal.
Query Syntax
:SBUS<n>:LIN:SOURce?
The :SBUS<n>:LIN:SOURce? query returns the current source for the LIN
signal.
Return Format
<source><NL>
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:MODE" on page 598
":SBUS<n>:LIN:TRIGger" on page 487