464 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
CAN Id specification is set by the :SBUS<n>:CAN:TRIGger:PATTern:ID
and:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE commands.
CAN Data specification is set by the :SBUS<n>:CAN:TRIGger:PATTern:DATA
command.
CAN Data Length Code is set by the
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth command.
Query Syntax
:SBUS<n>:CAN:TRIGger?
The :SBUS<n>:CAN:TRIGger? query returns the current CAN trigger on
condition.
Return Format
<condition><NL>
<condition> ::= {SOF | DATA | ERR | IDD | IDE | IDR | ALL | OVER | ACK}
Errors "- 241, Hardware missing" on page 753
See Also "Introduction to :SBUS<n> Commands" on page 447
":SBUS<n>:MODE" on page 451
":SBUS<n>:CAN:TRIGger:PATTern:DATA" on page 465
":SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth" on page 466
":SBUS<n>:CAN:TRIGger:PATTern:ID" on page 467
":SBUS<n>:CAN:TRIGger:PATTern:ID:MODE" on page 468
":SBUS<n>:CAN:SIGNal:DEFinition" on page 461
":SBUS<n>:CAN:SOURce" on page 462