232 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
10 :CHANnel<n> Commands
:CHANnel<n>:PROTection
(see page 790)
Command Syntax
:CHANnel<n>:PROTection[:CLEar]
<n> ::= 1 to (# analog channels) in NR1 format| 4}
With the 2000 X- Series oscilloscopes, the analog channel input impedance
is always 1 M
Ω, so automatic overvoltage protection is not necessary (as it
is for channels with 50
Ω input impedance). There are no protection
settings to clear, so the :CHANnel<n>:PROTection[:CLEar] command does
nothing.
Query Syntax
:CHANnel<n>:PROTection?
The :CHANnel<n>:PROTection query always returns NORM (normal).
Return Format
NORM<NL>
See Also "Introduction to :CHANnel<n> Commands" on page 218
":CHANnel<n>:COUPling" on page 221
":CHANnel<n>:PROBe" on page 227