4 Connect the power supply to the oscilloscope and to the multimeter,
using the BNC tee and cables.
5 Adjust the power supply output so that the multimeter reading
displays the first Power supply setting value in table 4-4.
Wait a few seconds for the measurment to settle.
6
Press the V2 softkey, then position the V2 cursor to the baseline.
The V value on the lower line of the display should be within the test limits
of table 4-4. If a result is not within the test limits, see "Troubleshooting the
Oscilloscope," on page 4- 32.
7
Continue checking the voltage measurement accuracy with the
remaining Power supply setting lines in table 4-4.
Table 4-4 Voltage Measurement Accuracy
Volts/Div setting Power supply setting Test limits
5 V/Div 35 V 34.04 V to 35.96 V
2 V/Div 14 V 13.616 V to 14.384 V
1 V/Div 7 V 6.808 V to 7.192 V
0.5 V/Div 3.5 V 3.404 V to 3.596 V
0.2 V/Div 1.4 V 1.3616 V to 1.4384 V
0.1 V/Div 700 mV 680.8 mV to 719.2 mV
50 mV/Div 350 mV 340.4 mV to 359.6 mV
20 mV/Div 140 mV 136.16 mV to 143.84 mV
10 mV/Div 70 mV 68.08 mV to 71.92 mV
5 mV/Div* 35 mV 33.66 mV to 36.34 mV
2 mV/Div* 14 mV 12.66 mV to 15.34 mV
*Full scale is defined as 56 mV on the 5 mV/div and 2 mV/div ranges.. Full scale on all
other ranges is defined as 8 divisions.
8 Disconnect the power supply from the oscilloscope, then repeat
steps 1 to 7 for channel 2.
Service
Verifying Oscilloscope Performance
4-9