To reduce the random noise on a signal
If the signal you are applying to the oscilloscope is noisy (figure 2-21), you
can set up the oscilloscope to reduce the noise on the waveform (figure
2-22). First, you stabilize the displayed waveform by removing the noise
from the trigger path. Second, you reduce the noise on the displayed
waveform.
1
Connect a noisy signal to the oscilloscope and press
Autoscale .
2 Obtain a stable display by removing the noise from trigger path;
press
Slope/Coupling , then select either the LF Reject softkey
or the
HF Reject softkey.
High frequency reject (HF Reject) adds a low pass filter with the 3 dB point at
50 kHz (see figure 2-19). You use HF reject to remove high frequency noise
such as AM or FM broadcast stations from the trigger path.
HF reject (trigger path)
3 dB down point
Pass
Band
dc
0 dB
50 kHz
Figure 2-19
Operating Your Oscilloscope
To reduce the random noise on a signal
2–30