Vista de pagina 94
Table 4-1
Recommended list of test equipment to service the oscilloscope
Equipment Critical specifications Recommended Model/Part Use
1
Signal generator 1 to 500 MHz at 200 mV
high stability timebase
Agilent 8656B Option 001 P
Digital multimeter 0.1 mV resolution, better than 0.01%
accuracy
Agilent 34401A P, A, T
Oscilloscope 100 MHz, 1 M input R Agilent 54600 P, T
Power meter and
Power sensor
1 to 500 MHz ±3% accuracy Agilent 436A and Agilent 8482A P
Power supply 14 mV to 35 Vdc, 0.1 mV resolution Agilent 6114A P
Pulse generator Rise time 700 ps Agilent 8131A A
Power splitter Outputs differ < 0.15 dB Agilent 11667B P
Shorting cap BNC Agilent 1250-0774 P
Time Mark
Generator
Stability 5 ppm after 30 minutes Tektronix TG501A and TM503B P
Dummy load
2
Compatible with power supply see note 2 below A
Adapter SMA (f) to BNC (m) Agilent 1250-1787 A
Adapter BNC (f-f) Agilent 1250-0080 P, A
Adapter BNC tee (m) (f) (f) Agilent 1250-0781 P, A
Adapter N (m) to BNC (f), Qty 3 Agilent 1250-0780 P
Adapter BNC (f) to dual banana (m) Agilent 1251-2277 P
Adapter Type N (m) to BNC (m) Agilent 1251-0082 P
Cable BNC, Qty 3 Agilent 10503A P, A
Cable BNC, 9 inches, Qty 2 Agilent 10502A P, A
Cable Type N (m) 24 inch Agilent 11500B P
Adapter
3
BNC (m) to dual banana post Agilent 10110B P
1
P = Use for Performance Verification.
A = Use for Adjustments.
T = Use for Troubleshooting.
2
See page 4-33 to construct your own dummy load.
3
Used for Option 005 only
Service
4-3