To trigger on all fields of the video signal
To quickly and easily view transitions between fields, or to find the amplitude
differences between the fields, use the All Fields trigger. The oscilloscope
will trigger on the first field it finds at the start of acquisition.
1
Select the TV display, TV as the trigger mode, and the appropriate TV
standard as described in the section, "To autoscale on a video signal."
2 Press Slope/Coupling in the TRIGGER section of the front
panel, then press the
Mode softkey until All Fields appears.
Triggering on All Fields
Figure 3-4
Using Option 005 Enhanced TV/Video Trigger
To trigger on all fields of the video signal
3–9