To reset the instrument setup
1 To reset the instrument to the default factory-preset configuration,
press
Setup .
2
Press the Default Setup softkey.
3
To reset the instrument to the configuration that was present before
pressing
Autoscale, press the Undo Autoscale softkey.
Table 2-1
Default Setup configuration settings
Configuration Item Setting
Cursors Cursors off; time readout is selected; all cursors are set to time/voltage zero.
Trace memories Both trace memory 1 and 2 are off; trace 1 memory is selected.
Setup memories Setup memories are off; setup memory 1 is selected.
Graticule Grid set to Full
Autostore Off
Time base Time reference center; main, not delayed sweep; main and delay value 0; 100 µs/div
main time base; sample rate is 5 MSa/s.
Display Vectors On, Display Mode Normal.
Channels Channel 1 on, Position 0 V, Volts/Div 100 mV.
Trigger Mode Auto Level, Coupling DC, Reject Off, Noise Reject Off.
Trigger Condition Rising edge of channel 1
See Also "To Clear Error Messages" in the troubleshooting section of Service
chapter 4.
Operating Your Oscilloscope
To reset the instrument setup
2–35